Page 156
3.
Mount the Dial Indicator Mount and Precision Dial Indicator.
4.
Adjust the tip on the Precision Dial Indicator to strike the equator of the ball. Make sure the
Indicator is not pegged (i.e. the small needle on the gauge, near the centre of the Dial Indicator,
is around the halfway mark).
Note:
The Precision Dial Indicator is very sensitive and care should be taken not to lean or touch any
part of the system while adjusting the Indicator or the X,Y Centring Ball tool.
5.
Zero the Precision Dial Indicator.
Note:
For fine measurements, the Precision Dial Indicator may not be accurate until after the first
movement.
6.
Use the
Rotation
setting in
X-Ray and Motor Controller
to rotate the ball 360° (180° at a time).
Watch as the ball is rotating to ensure that the tip of the Precision Dial Indicator is touching the
ball at all points of rotation and that it is not pegged. Re-zero the Dial Indicator, if necessary.
Centre the Ball:
7.
Use the
Rotation
setting to rotate the ball 180°.
8.
Watch the Precision Dial Indicator as the ball rotates. Note the reading of the Precision Dial
Indicator after 180° of rotation (e.g. .0004") in the Worksheet.
The objective is to adjust the ball so that, after the adjustment, the Dial Indicator reads half of
the original reading (e.g. .0002").
To do this:
9.
Use the tool’s adjustment key to carefully turn the 2 adjustment screws that are aligned with
the Dial Indicator - turn the screws until the Dial Indicator reads half of the original reading.
NOTICE
Equipment
Damage
Possible
The tool is extremely delicate. Fine
adjustments are required. DO NOT
overtighten the screws. One screw
turns clockwise and the other turns
counter clockwise. Turn the screws in
tandem using as little torque as
possible (i.e. 5° at a time)
10. Once adjusted, rotate the ball 180°
and check the Dial Indicator reading. It
should match the reading taken after
the adjustment in the previous step
(e.g. .0002").
11. Re-zero the Dial Indicator.
12. Use the rotation setting to rotate the
ball 360° (180° at a time). Watch as
the ball is rotating to ensure that the
tip of the Precision Dial Indicator is
touching the ball at all points of
rotation and that it is not pegged. Re-
zero the Dial Indicator, if necessary.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...