Page 62
Section 6.2
Remove / Replace FRUs
6.2.1
Comair Blower and Hose Removal / Replacement
The blower is located behind the left panel assembly.
TOOL LIST
Standard toolkit
SUPPLIES
Lockwashers
SAFETY CONSIDERATIONS
X-ray shielding contains lead and lead compounds, chemicals known (to the State of California) to
cause cancer and other reproductive harm. Wear gloves when handling shielding and wash hands
after handling.
PRECONDITIONS
•
Machine locked out See 6.1.1 for details..
•
Left access panel removed. See 6.1.2 for details.
REMOVE
1.
Disconnect power cable from blower.
2.
Using slotted screwdriver loosen clamp
securing ventilator hose to blower.
3.
Using 5/32” hex key, remove two screws and
lockwashers securing blower mounting bracket
to unit base. Remove blower, discard
lockwashers.
4.
Using Phillips screwdriver remove two screws
and flatwashers securing mounting bracket to
blower.
5.
Using 9/64” hex key, remove blower manifold
from blower.
Manifold
5/32"
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...