6.2.4
Interlock Removal / Replacement
INTRODUCTION
The unit has three interlocks: on right and left panel and on the sliding door.
The left and right interlocks are identical, and this procedure applies to both. See Section 6.2.19 for
details on the door interlock procedure.
TOOL LIST
Standard toolkit
SUPPLIES
Lockwashers
SAFETY CONSIDERATIONS
X-ray shielding contains lead and lead compounds, chemicals known (to the State of California) to
cause cancer and other reproductive harm. Wear gloves when handling shielding and wash hands
after handling.
PRECONDITIONS
•
Machine locked out. See 6.1.1 for details.
•
Left or right access panel removed, as required. See 6.1.2 and 6.1.3 for details.
REMOVE
1.
Using 9/64” hex key, remove two screws and
internal tooth lockwashers securing interlock to
bracket. Discard lockwashers.