GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 70
6.2.6
Temperature Controller Removal / Replacement
INTRODUCTION
The temperature controller is the metal blue box,
usually located in storage area.
TOOL LIST
None
SUPPLIES
None
SAFETY CONSIDERATIONS
n/a
PRECONDITIONS
Machine locked out. See 6.1.1 for details.
REMOVE
1.
Disconnect three cables from back of
temperature controller.
REPLACE
Reverse of removal.
FOLLOW-ON PROCEDURES
•
Temperature controller should be programmed
prior to shipment to site.
•
Target temperature is 12.98. See set-up
procedure, below.
Target temperature
- fixed
Actual temperature
- fluctuates