43. Threshold the image. Slide the
Window
slide bar to the left until it reads 0 and slide the
Level
slide bar until the ball appears with the maximum amount of square edges.
The
Window
and
Level
values control the image’s contrast and brightness. These values are
adjusted using either the slide bars at the bottom of the
MicroView
window, or by right-clicking
on either slide bar to display a pop-up window where you enter a specific value.
44. Zoom in on the outer right most side of the ball.
45. Place the mouse in the middle of the gray band that appears to the right of the black area.
Make note of the x coordinate in the Worksheet.
46. Zoom out and move to the left outer most side of the ball. Zoom in on the edge.
Note:
DO NOT
re-adjust Window and Level slide bars.
47. Place the mouse in the middle of the gray
band that appears to the left of the black
area. Make note of the x coordinate in the
Worksheet.
48. Add the 2 x values together and divide by 2. This value is the ball’s geometric centre. Make
note of this value in the Worksheet.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...