Page 92
6.2.17
Fan 2.0625 (Below X-Ray Source)
INTRODUCTION
This fan is located below and to the rear of the x-ray
source. It can be removed from the left-hand side
and through the door with the case in place.
TOOL LIST
Standard toolkit
SUPPLIES
Mirror
SAFETY CONSIDERATIONS
None
PRECONDITIONS
•
Machine locked out. See 6.1.1 for details.
•
Left access panel removed. See 6.1.2 for details.
•
Left x-ray shield removed. See 6.1.5 for details.
REMOVE
1.
Using 9/64” hex key, remove four hex screws and lockwashers and flatwashers securing fan
to adjusting plate. Fan will pull down through the hole once the wires are free.
2.
Cut cable ties securing fan wires
3.
Using mirror and small slotted screwdriver, locate and remove
fan wires from terminal. Note the wire location.
REPLACE
1.
Insert and secure fan into place and feed wires out to rear of
unit. Fan must be blowing down and away from x-ray source.
2.
Attach fan wires and secure to terminal block with slotted
screwdriver.
FOLLOW-ON PROCEDURES
None
Fan, 2.0625
Fan wires
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...