GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 71
6.2.6.1
Set-Up Temperature Controller
1.
Plug in and run the Temperature Controller.
2.
Press the Green button on the left of the control panel to scroll to the first setting.
Pcnt
should
appear in green at the bottom of the display screen.
3.
Ignore the
Pcnt
setting and press the green button again until
PBl
appears in green on the
bottom of the display.
4.
Enter the values below as you scroll through the settings, pressing the green button after each
entry. Use the up and down arrow buttons on the right of the panel to enter values.
5.
Press the middle button
°°
upon completion.
6.
Enter the temperature setting of 12.98 in the bottom part of the display.
Setting
Value
PB1
2.99
1t1
7.00
dE1
0.98
Pb2
6.26
1t2
7.00
dE2
0.98
AtLl
100.0
AtL2
100.0
lLSP
0.00
lLA
50%
lLb
50%
A3Lo
1.00
A3h1
22.00
A-77
AUTO
Table 6-1 Settings