Page 238
8.2.17
Equalization Bath
1.
Inspect the
equalization bath
and specimen
holder and check
for any cracking or
discolouration.
8.2.18
Temperature Controller
1.
Verify that the system is
running at approximately
12.98.
The number on the bottom
LED display (green) on the
temperature controller
should read 12.98. The
number displayed in the top
LED display (red) should be
approximately 13.
Note:
This is a resistance
number and not a degrees
setting.