6.2.13
X-Ray Source Removal / Replacement
INTRODUCTION
The x-ray source is located under a shielded
aluminum box at the left-side of the unit. It is
attached to an adjusting plate by four screws.
TOOL LIST
Standard toolkit
SUPPLIES
Cable ties
Lockwashers
SAFETY CONSIDERATIONS
X-ray shielding contains lead and lead compounds,
chemicals known (to the State of California) to cause
cancer and other reproductive harm. Wear gloves when handling shielding and wash hands after
handling.
PRECONDITIONS
•
Machine locked out. See 6.1.1 for details.
•
Left and right access panels, case removed. See 6.1.2, 6.1.3 and 6.1.4 for details.
•
Equalization bath bracket removed. See 6.2.7 for details
•
Remove filter wheel. See 6.2.9 for details.
Note:
The x-ray source may be removed without taking off the right cover or case. In order to do this,
a shortened hex key will be needed to access the 4 screws on the top of the X-ray source
shield.
REMOVE
1.
Loosen and remove rear source shielding using thumbscrews. Set aside.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...