Page 109
23. Access
Calibration Tools
and click the
Transform
tab.
24. Complete the Example VFF File field using
Browse
to find the
corrected_grid.vff
file.
25. Ensure the Source/Targets File field is set to
C:\translations
26. Leave the Tilt values field as zeros.
Note:
If not zeros, click the
Calculators
tab and press
Reset
under Tilt Calculator, then press
Update
.
Return to the
Transform
tab and press the
Update
button.
27. Ensure 1x1 appears in the Transform bin Mode field and Acquired Image bin Mode field.
28. Press the
Create D:\Transform1 and Transform1.dat
button at the bottom of the window.
29. Use
Windows Explorer
to rename the
Transform1
file and the
Transform1.dat
files to
Transform01
and
Transform01.dat
.
30. Access
Calibration Tools
and click the
Unwarp
tab and complete the fields as follows:
VFF to unwarp:
D:\calib\corrected_grid
Transform grid:
D:\Transform01
31. Click the
Unwarp
button.
32. Call the file
D:\zz
when
prompted to name the file,
then click
OK
.
The zz file will be used as a
temporary file for unwarping.
Continue to overwrite this file
whenever unwarping.
33. Double-click
zz
in
Windows
Explorer
to load the
unwarped corrected_grid file
into
MicroView
.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...