Page 208
6.
Find the motor limit switch (this ensures that the limit switches will never be triggered):
-
select
Magnification
from the drop down menu in the Motor field
-
press the
Left (CW) Linear Home
button to move the stage toward the Source
-
enter
2
in the mm field, click on the
right or CCW
radio button, then press the
Linear Stage
Motion
button to offset the "home" position by 2 mm.
7.
Click the
Start Shutter
button.
8.
Click the
Turn X-Ray
On button.
9.
Click the
Fluoro
button, then:
-
select
OK
when the warning pop-up appears.
-
select
OK
when the text concerning placement of the specimen appears.