Page 234
8.2.5
Limit Switches on Magnification Stage
1.
Visually inspect each limit switch and verify that
nothing is blocking the sensor of the limit switch.
2.
Access the
X-ray and Motor Controller
software and verify that sensor indicates it is
clear.
3.
Select
Magnification
from the drop down menu
in the Motor field. Press the
Left (CW) Linear
Home
button to move the stage toward the
detector. Verify that the software is registering
the trigger - displays 'Left (CW) is ON' in the
Limit Switch Status.
4.
Press the
Right (CCW) Linear Home
button to
move the stage toward the source. Verify that 'Right (CCW) is ON' is displayed in the Limit
Switch Status.