6.1.5
X-Ray Shield Removal / Replacement
INTRODUCTION
Shields must be removed to access the detector,
source and stage. Typically, back shield is
removed first.
TOOL LIST
Standard toolkit
SUPPLIES
None
SAFETY CONSIDERATIONS
X-ray shielding contains lead and lead
compounds, chemicals known (to the State of
California) to cause cancer and other reproductive
harm. Wear gloves when handling shielding and
wash hands after handling.
PRECONDITIONS
•
Machine locked out. See 6.1.1 for details.
•
Left and right access panels, case removed. See 6.1.2, 6.1.3 and 6.1.4 for details.
REMOVE
1.
Using 9/64” hex key, remove four screws and
flatwashers securing back shield to right and
left shields. Set shield aside.
Rear shield
Right shield
Left shield
9/64"