Page 181
7.4.3
Calculate Central Slice of X-Ray Source
This procedure calculates the Central Slice of the X-ray. With the X,Y Centring Ball tool still
mounted and the system warmed up:
1.
Move the stage back toward the detector using the
X-Ray and Motor Controller
.
2.
Elevate and rotate the tool until both bebes appear in the
centre of the screen in
Fluoro
.
3.
Capture the image and name it
CS1
.
4.
Move the tool 120mm toward the X-Ray Source and capture
an image. Name the image
CS2
.
5.
Correct the CS1 image, unwarp it and load it into
MicroView
.
6.
Determine the y coordinates of both bebes using the Region
Grow tool in
MicroView
, as follows:
a.
Zoom in on the top bebe and find the centre of the bebe. Adjust the Window and Level to
make the bebe as round as possible. Record Window and Level in Worksheet.
b.
In
MicroView
, choose
Tools>Region Grow.
c.
Select
Lower
in the Threshold field.
Содержание Healthcare eXplore Locus SP