Page 219
h.
Make a note of this value in the Worksheet. This is the ideal Centre of Rotation at the
Central Slice.
36. Click the
Advanced
button in the
Reconstruction Utility
window, and choose the
Reconstruction
tab.
37. De-select the Short Scan field and click
OK
.
38. Click the
Recon
button in the
Reconstruction Utility
window.
39. Name the file
1x1_wire_cs_360
(or
2x2_wire_cs_360
or
4x4_wire_cs_360
depending on
which verification scans are being performed)
40. Confirm the wires are still round in
MicroView
. The wires should appear even rounder. Exit
MicroView
.
Note:
If the wires appear hollow (like a ring) in the 360 degree reconstruction, then the Centre of
Rotation needs to be adjusted as described above. Also, verify that the HU calibration values
are set to -1.
This process has determined the ideal Centre of Rotation of the wire phantom at the Central Slice.
The following steps check the Centre of Rotation of the image at either end of the wire phantom.
VERIFY CENTRE OF ROTATION AT EACH END (FOR 1X1 MODE
ONLY
):
1.
Change the Z-Origin field on the
Reconstruction Utility
window to
200
. This represents one
end of the wire phantom.
2.
Click the
Advanced
button, then choose the
Reconstruction
tab.
3.
Re-select Short Scan, then click
OK
.
4.
Click the
Recon
button. Name the file
1x1wire_-end_ss
5.
View the image in
MicroView
. The wires should appear as round as possible.
If the wires are not round, the Centre of Rotation needs to be adjusted as described above.
6.
Upon completion, make note of the difference between this Centre of Rotation and the ideal
Centre of Rotation at the Central Slice, in the Worksheet.
Note:
The value of the Centre of Rotation at either end of the wire phantom may only differ from the
ideal Centre of Rotation at the Central Slice by 1.5 pixels. The Short Scan option in
Reconstruction Utility
must be used for this test to be valid.
7.
Change the centre of rotation back to the ideal found at the Central Slice and repeat the
reconstruction with all 360 views (de-select the Short Scan field). Name the file
1x1wire_-
end_360
. Verify that the wires appear round.
8.
Move to the other end of the wire phantom by changing the Z-Origin to
2600
. Re-select the
Short Scan field. Name the file
1end_ss
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...