Note:
Details concerning the following steps are found in the
eXplore Locus SP User Guide
. Refer to it
for particulars.
3.
Warm up the X-Ray Source.
4.
Once warmed up, launch
eXplore Scan Manager
.
5.
Click
Scan New Subject
and name the subject
1x1
(or
2x2
or
4x4
depending on which
verification scans are being performed).
6.
Enter
1x1wire
in the Scan Description field (or
2x2wire
or
4x4wire
depending on which
verification scans are being performed).
7.
The
Scan Control
window opens. Select
1x1wire
from the Protocol drop-down list (or
2x2wire
or
4x4wire
depending on which verification scans are being performed).
8.
Select the
Scan bright and dark first
,
Recon after scan
and
Home all stages
checkboxes under
Options.
9.
Start
Fluoro
.
10. Click
Menu
and choose
Position Stages
from the
drop-down list.
11. Adjust the elevation until the wires of the center
section of the wire phantom extend from the top to
the bottom of the screen.
12. Exit
Fluoro
.
13. Press the
Start Scan
button (the scan takes
approximately 45 minutes) and follow the prompts
to place and remove the specimen from the
scanner for the bright and dark scans.
The
eXplore Reconstruction Utility
launches
automatically when the scan is complete.
14. Click the
Set to Central Slice
button.
15. Click the
Thumbnail
button.
16. Change the Z-Size field to
5
.
17. Press the
Recon
button on the
Reconstruction
Utility
window.
18. View the image in
MicroView
. Draw a line from the wire on the left side of the screen to the
wire on the top using the 1 and 2 keys, and note the angle displayed in the top left of the screen
in the Worksheet.
19. Exit
MicroView
.
20. Press the
Advanced
button on the
Reconstruction Utility
window and select the
Reconstruction
tab.
21. Enter the approximate angle of the 4 wires noted above in the Angle field.
22. Record the existing air, water and HU calibration values in the Worksheet. Enter
-1
(minus one)
values for each field.
23. Select
Short Scan
. This looks at the first 200 views instead of all 360 views. It will emphasize
any errors in calibration. Click OK.
24. Press the
Recon
button in the
Reconstruction Utility
window.
25. View the image in
MicroView
upon completion. The 4 wires should appear "square" to the
screen.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...