2.
Using 9/64“hex key, remove two screws
securing interlock to bracket.
3.
Using Phillips screwdriver, open access cover
and remove wires.
4.
Using 3/8” socket, remove the four tamper-
proof screws securing the switch mounting
block to the case.
5.
Using a narrow slotted screwdriver, loosen the
cover on back of switch.
6.
Using slotted screwdriver, loosen two terminals
and remove wires from terminal and through
holes in block.
7.
Remove switch from mounting block to replace.
8.
Disconnect wires into switch.