GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 40
-
right-click the icon and select
System Idle
from the list of options
-
right-click the icon again and select the
<=10 Minutes to Scan
option
-
once the cycle has finished (approximately 14 minutes), right-click on the icon again and
select
Exit
.
The
System Warm-Up
application can be re-initialized by double-clicking on WarmupApp.exe
in C:\Program Files\GE Medical Systems\eXplore Utilities\
5.
Access the
X-Ray and Motor Controller
software.
6.
Ensure the fields below are set to the following settings:
Camera:
Thomson 8M
Bin Mode:
1 x 1
Exposure Time:
3000 ms
kV:
80
µA:
80
7.
Ensure the X-Ray switch on the unit is ON and the red
Reset
button is not lit.
8.
Click on the
Start Shutter
button on the
X-Ray and Motor Controller
screen.
9.
Click on the
Turn X-Ray On
button.
10. Click on the
Fluoro
button, then:
-
select
OK
when the warning pop-up appears.