6.2.8
Magnification Stage Limit Switches Removal / Replacement
INTRODUCTION
The magnification stage is equipped with two limit
switches, located along the outside of the limit
switch bracket.
TOOL LIST
Standard toolkit
SUPPLIES
None
SAFETY CONSIDERATIONS
X-ray shielding contains lead and lead
compounds, chemicals known (to the State of
California) to cause cancer and other reproductive
harm. Wear gloves when handling shielding and
wash hands after handling.
PRECONDITIONS
•
Machine locked out. See 6.1.1 for details.
•
Door open.
REMOVE
1.
Unplug cable connector from limit switches.
2.
Using small Phillips screwdriver, remove screw and lockwashers securing limit switch to
bracket.
REPLACE
Reverse of removal.
FOLLOW-ON PROCEDURES
Run the X-Ray Motor Controller application and verify that the limit switches are communicating
with the software and that the stage/slide does not crash.
Magnification stage limit switches
Shown with cover
removed for clarity