10. Change the centre of rotation back to the ideal found at the Central Slice and repeat the
reconstruction with all 360 views (de-select the Short Scan field). Name the file
1end_360
. Verify that the wires appear round.
11. Once successful, click the
Set to Central Slice
button on the
Reconstruction Utility
window.
12. Click the
Advanced
button and choose the
Reconstruction
tab.
13. Change the Centre of Rotation field back to the value of the ideal Centre of Rotation at the
Central Slice, then click
OK
.
14. Click the
Recon
button on the
Reconstruction Utility
window.
15. View the image in
MicroView
and zoom in on the wire that appears least round to be
measured.
16. Select a region of interest within the phantom housing near the wire being measured. Press
m
to calculate the mean value of the plastic area selected. Record in Worksheet.