Page 153
If the values
ARE NOT
within 2 pixels of each other:
31. Place shims under the elevation rail (loosen stage bolts, etc.).
Note:
One pixel is approximately .002" in shims. If the x value of the top bebe in
Fluoro
is less than the x
value of the bottom bebe, the shim goes on the bottom of the stage. If the x value of the top bebe
is higher, the shim goes at the top of the stage.
The shims are placed close to the
mounting bolts. Larger adjustments
(more than 0.004") will require shims at
both the top
and
the bottom to allow the
stage to tip once it is tightened (e.g.
large shim on the bottom and small shim
on the top). This is because flat surfaces
are bolted together.
NOTICE
Equipment
Damage
Possible
Failure to provide a tip relief will cause the rail
assembly to bend when the bolts are tightened.
There should be at least 0.001" clearance at the tip
end
.
If you have to shim the stage, re-check the stage
tilt mechanically using the 9" Square.
Repeat steps 18 to 29 until the values are
or - 2 pixels.
Shims
Stage Outer
Bracket
tip relief
no tip relief
shims
0.001"
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...