GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 182
d.
Place the mouse over the bebe and press the Spacebar (the bebe’s solid centre turns
green).
e.
Record the centroid value (y coordinate only). If desired, press the
View Centroid
button
to view the calculated actual centre (see below). You may have to Pan and Zoom to see
the cross hairs.
7.
Make note of the CS1 file’s two y coordinates (top and bottom bebe) in the Worksheet.
8.
Correct the CS2 image, unwarp it and load it into
MicroView
. Determine the y coordinates of
both bebes using MicroView’s Region Grow tool.
9.
Make note of the CS2 file’s two y coordinates (top and bottom bebe) in the Worksheet.
10. Access
Calibration Tools
and select the
Calculators
tab.
11. Enter the 4 following values under Central Slice Calculator in the order below:
y1a pos:
enter the CS1 file’s top bebe’s y coordinate
y2a pos:
enter the CS1 file’s bottom bebe’s y coordinate
y1b pos:
enter the CS2 file’s top bebe’s y coordinate
y2b pos:
enter the CS2 file’s bottom bebe’s y coordinate
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...