GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 122
18. Once the ball has been centred, remove the Dial Indicator Mount and Precision Dial Indicator.
19. Select
Magnification
in the Motor field in the
X-Ray and Motor Controller
window and move
the ball 70mm toward the detector. This should be 2mm from the limit switch.
20. Warm up the x-rays without using the
System Warm-up
application to avoid moving the stage.
To do this, use the
X-Ray and Motor Controller
program and do the following:
a.
Set the kV field to 80
b.
Set the µA field to 10
c.
Press the
Turn X-Rays
on button
d.
Wait 2 minutes
e.
Press the
Turn X-Rays
off button
f.
Set the kV and µA field to the next value in list below, and repeat steps c to e:
- wait 2 minutes @ 80kV, 45µA
- wait 10 minutes @ 80kV, 80µA
21. Capture the image in
Fluoro
and name it
capture
.
22. Correct and unwarp the captured image using
Calibration Tools
.
Find the Geometric Centre of the Ball at Detector End:
Note:
To find the geometric centre of the ball, determine the x value on the right and left side of the ball.
Add the values together and divide by 2. To find the geometric centre as follows.
23. Use the Pan and Zoom features in
MicroView
to get the ball in the desired area of the screen.
Note:
To Pan, move the mouse while pressing Shift plus the left mouse button simultaneously. To
Zoom, move the mouse up (to magnify) or down (to zoom out) while pressing the right mouse
button.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...