Page 163
34. View the cropped area in
MicroView
upon
completion.
35. Zoom in on individual wires. All 4 wires should be as
round as possible.
If all of the wires are not round, the Centre of Rotation needs to be adjusted as follows:
a.
Exit
MicroView
.
b.
Click the
Advanced
button in the
Reconstruction Utility
window.
c.
Click the
Reconstruction
tab.
d.
Change the Centre of Rotation field. Make only small, incremental changes to this number
(e.g. change 1150 to 1150.5) then click
OK
.
e.
Click the
Recon
button in the
Reconstruction Utility
window.
f.
View the image in
MicroView
upon completion. If the image is improved, continue to
change the Centre of Rotation in the same direction (e.g. change 1150.5 to 1151). If the
image is worse, change the Centre of Rotation in the opposite direction (e.g. change
1150.5 to 1149.5).
g.
Continue to adjust the Centre of Rotation until the wires are as round as possible (although
all 4 wires are unlikely to be equally round).
h.
Make a note of this value in the Worksheet. This is the ideal Centre of Rotation at the
Central Slice.
If the wires appear square, ensure that the interpolation is on:
a.
Go to Edit->Application Settings
b.
Ensure a checkmark appears next to “Use Image Interpolation”.
Round Wire
Skewed Wire
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...