Page 177
The x value must eventually equal the centre of rotation in the registry (within 5 pixels). Open
Calibration Tools
(located in C:\Program Files\GE Medical Systems\eXplore
Utilities\Calibration\). Click on
Calib. Settings
tab. Click
File
, and choose
Read Registry Values
from the drop-down menu. The Centre of Rotation equals the value in Centre of Rotation1.
This positioning puts the ball in the centre of the detector.
To reach the correct x value:
a.
Turn off the X-Ray Source.
b.
Loosen the 2 screws on
the top of the Source and
move the Source using the
Source adjustment wheel.
This moves the Source
along the x-axis as shown
in the diagram.
c.
Repeat steps 34 to 48 until
the x value equals the
same value that appears in
the registry (within 5
pixels). Use the Worksheet
to record values.
Once the ball is in the centre of the detector:
49. Tighten the screws on the top of the X-Ray Source.
50. Tightening the screws may change the value of the ball centre. Turn X-rays back on, and verify
the setting by repeating steps 34 to 48.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...