11. Repeat the above step in the black area of the screen. Note this value in the Worksheet.
12. Add the two mean values determined in the above 2 steps.
13. Divide this value by 2 to determine the Half Maximum value. Note this value in the Worksheet.
14. In
MicroView
, slide the
Window
slide bar left until it reads 0.
15. Double-click on the W and L parameters in the lower right of the screen. Enter the Half
Maximum value calculated in step 13 in the Level field.