Page 66
6.2.3
Left Comair Fan Removal / Replacement
INTRODUCTION
The left fan is located on the rear panel, and blows
air out of the unit.
TOOL LIST
Standard toolkit
SUPPLIES
None
SAFETY CONSIDERATIONS
X-ray shielding contains lead and lead compounds,
chemicals known (to the State of California) to cause
cancer and other reproductive harm. Wear gloves
when handling shielding and wash hands after handling.
PRECONDITIONS
•
Machine locked out. See 6.1.1 for details.
•
Remove left access panel. See 6.1.2 for details.
REMOVE
1.
Using Phillips screwdriver, remove four
screws, flatwashers, and finger guard
securing fan to rear panel.
2.
Using slotted screwdriver, disconnect
ground wire, and unplug power wires
from fan.
REPLACE
1.
Reverse of removal.
2.
Ensure that air flows out of unit.
3.
Power wires to fan are not polarized.
FOLLOW-ON PROCEDURES
Observe that the fans operate with the
system power on.
Right
fan
Left
fan
Left fan is
unshielded
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...