GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 81
B) REMOVE ENTIRE ASSEMBLY, THEN SEPARATE
1.
Disconnect cables (4) to specimen slide.
2.
Using 4mm hex key, remove the stage
retaining screws.
3.
Remove entire rotary stage and specimen
slide assembly.
4.
Separate the rotary stage from specimen
slide:
Using 4mm hex key ball driver and 8mm socket, remove four screws and nuts securing
specimen slide to angle plate.
5.
Using 2.5mm hex key remove two screws and
lockwashers securing limit finger to specimen
slide.
6.
Using 3/32” hex key remove limit switch bracket
from specimen slide.
4mm
Angle plate
4mm
8mm
2.5mm