GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 217
17. Press the
Recon
button on the
Reconstruction Utility
window.
18. View the image in
MicroView
. Draw a line from the wire on the left side of the screen to the
wire on the top using the 1 and 2 keys, and note the angle displayed in the top left of the screen
in the Worksheet.
19. Exit
MicroView
.
20. Press the
Advanced
button on the
Reconstruction Utility
window and select the
Reconstruction
tab.
21. Enter the approximate angle of the 4 wires noted above in the Angle field.
22. Record the existing air, water and HU calibration values in the Worksheet. Enter
-1
(minus one)
values for each field.
23. Select
Short Scan
. This looks at the first 200 views instead of all 360 views. It will emphasize
any errors in calibration. Click OK.
24. Press the
Recon
button in the
Reconstruction Utility
window.
25. View the image in
MicroView
upon completion. The 4 wires should appear "square" to the
screen.
26. Select a region of interest that encompasses all 4 wires using the 7 and 8 keys in
MicroView
.
27. Select File>Save crop coordinates.
28. Exit
MicroView
.
29. Click the
Load Crop
button on the
Reconstruction Utility
window.
30. Press the
Recon
button. View the image in
MicroView
and verify the image is cropped
properly, then exit
MicroView
.
31. Change the Z-Size field on the Reconstruction Utility screen to
1
.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...