Page 151
28. Double-click
on the W and
L parameters
in the lower
right of the
screen.
Change the
Update mode
to
Delayed
updates
. This
only needs to
be changed
once.
29. Make a note of the x value of the centre of the bebe using
MicroView’s
Region Grow tool:
-
In
MicroView
, choose
Tools>Region Grow.
-
Select
Lower
in the Threshold field.
-
Place the mouse over the bebe and press the Spacebar (the bebe’s solid centre turns
green).
-
Record the centroid value (x & y coordinates) in the Worksheet. If desired, press the
View
Centroid
button to view the calculated actual centre. You may have to Pan and Zoom to
see the cross hairs.
30. Repeat the steps above (Correct the Image, and Unwarp the Image) for the
shim2
file. Use the
Worksheet and make a note of the x value of the centre of the bebe using
MicroView’s
Region
Grow tool.