GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 202
23. Move the mouse to the right
line (side) and press
2
at the
Full Width Half Maximum
value point to draw a line from
one side to the other.
24. Make note of the line length
in pixels, in the Worksheet.
25. In
eXplore Scan Manager
,
navigate to the 1x1wire scan.
Note the resolution value
displayed on the right-hand
side of the window, in the
Worksheet.
26. Divide .025mm (which is the
size of the wire in the wire
phantom) by the resolution
value and then add 2 to the
result. This value is the target
value, in pixels, of the wire
scan.
Example:
.025 ÷ .007634 = 3.274 (target value)
3.274 + 2 = 5.274 (maximum value)
Record values in the Worksheet.
27. Verify that the line length calculated in step 25 is less than the maximum value calculated in
step 27. If not, calibration will need to be restarted.
28. Repeat steps 15 to 28 for the remaining 3 wires. Use the Worksheet to record values, as
appropriate.
29. Exit the plot, then exit
MicroView
.
Upon completion:
In the
Reconstruction Utility
, press the
Advanced
button. Go to the Recon tab. Re-enter the noted
air, water and bone HU calibration values into each field for all three scans (1x1, 2x2 and 4x4).
Inform the customer that they may need to re-calibrate their HU values.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...