Page 105
Capture an Image:
21. Capture an image in
Fluoro
as follows:
-
click
Menu
at the top of the
Fluoro
screen and choose
Capture
from the drop-down menu
-
once the Capture pop-up is displayed, enter
1
in the Number of Frames to Capture field
-
enter
10
in the Number of Frames to Average
field
-
name the file
grid_1
in the Name of Series
field
-
save the file to the directory: D:\calib
-
click the
Capture
button.
The progress of capture is shown in the Fluoro’s
title bar as
[saving...]
. Wait until the image is
captured.
22. Turn off the X-Ray Source.
23. Capture a dark image in
Fluoro
, and call this
image
dark
.
24. Remove the bebe grid.
25. Turn the X-rays back on and wait a few minutes for the Source to stabilize.
26. Capture a bright image in Fluoro and call this image
bright.
Keep X-Ray Source warm by leaving it on.