Page 75
6.2.9
Filter Wheel Assembly Removal / Replacement
INTRODUCTION
The filter wheel assembly is secured to the front of
the x-ray source by a cover.
TOOL LIST
Standard toolkit
SUPPLIES
None
SAFETY CONSIDERATIONS
None
PRECONDITIONS
•
Machine locked out. See 6.1.1 for details.
•
Door open.
REMOVE
1.
Rotate shutter to better access filter wheel.
2.
Using 1/16” hex key, remove three hex screws securing filter wheel retainer to filter wheel
mount base. Support wheel while removing last screw.
REPLACE
1.
Ensure filter numbers on front of wheel are visible.
2.
Place filter wheel onto shaft.
3.
Secure filter wheel retainer with three screws and 1/16” hex key. Ensure wheel can rotate
easily.
FOLLOW-ON PROCEDURES
None
Filter wheel
Shutter
Retainer