Page 183
12. Press the
Update
button under Central Slice Calculator.
13. Make note of the Central Slice value (e.g. 1453.25) in the Worksheet.
7.4.4
Verify Stage Alignment
VERIFY STAGE TILT (FRONT-TO-BACK MOVEMENT):
1.
Turn off x-rays and adjust the X,Y Centring Ball tool so that the bebes are facing the detector.
2.
Turn x-rays back on and wait a few minutes for the Source to stabilize.
3.
Move the stage (using
X-Ray and Motor Controller
) 120mm back toward the detector.
4.
Access
Fluoro
. and press
m
to display the mouse position window.
5.
Rotate the tool until the top bebe appears in the centre of the x axis of the
screen.
6.
Adjust the elevation using the
X-Ray and Motor Controller
program so
that the top bebe moves downward and appears within 200 pixels of the
bottom of the screen.
7.
Capture an image in
Fluoro
and name it
shim1.
8.
Adjust the elevation, so that the top bebe moves upward and appears within 200 pixels of the
top of the screen (approximately 50mm).
9.
Capture an image in
Fluoro
and name it
shim2
.
10. Correct the Shim1 file and unwarp it (using Transform1).
11. Load the corrected and unwarped file into
MicroView
.
12. Use the Worksheet and make a note of the x value of the centre of the bebe using
MicroView’s
Region Grow tool.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...