Page 77
6.2.11
Rotary Stage Removal / Replacement
INTRODUCTION
The rotary stage is secured to the gear box slide
spacer with four screws and two dowels. It pivots
on the dowels for calibration and is secured by the
screws.
TOOL LIST
Standard toolkit
SUPPLIES
Cable ties
Lockwashers
SAFETY CONSIDERATIONS
X-ray shielding contains lead and lead
compounds, chemicals known (to the State of
California) to cause cancer and other reproductive harm. Wear gloves when handling shielding and
wash hands after handling.
PRECONDITIONS
•
Machine locked out. See 6.1.1 for details.
•
Left access panel removed. See 6.1.2 for details.
REMOVE
1.
Using 7/64” hex key, remove equalization support bracket from top of stage.
2.
Use 9/64” t-handle allen wrench, remove screw from front of bracket.
3.
Cut cable ties securing two cables to rotary stage. Disconnect cables to rotary stage.
4.
Using 5/32” hex key, remove four screws and
lockwashers from back of rotary stage.
Discard lockwashers.
Rotary stage
disconnect cables
5/32"
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...