Page 132
17. Change the Z-Size value to
1
.
18. Click on the
Recon
button.
19. View the reconstructed image in
MicroView
.
20. Change the
MicroView
settings to display measurements in pixels:
-
select Edit>Application Settings
-
select the Miscellaneous tab
-
de-select
show measurements in mm
21.
Find the widest point
by plotting a line from one side of the plastic dowel to the other:
-
zoom in on the left side, place your mouse on the furthest edge of the plastic dowel,
and press
1
-
pan over to the right side’s furthest edge and press
2
(
you may need to press 2 repeatedly
at different points until you find the widest point or the maximum number of pixels
)
22. Determine the maximum number of pixels from one side of the plastic dowel to the other
(diameter) and note value in the Worksheet.
23. Calculate the measured pixel size using the following equation (and record in Worksheet):
measured pixel size = Measured Plastic dowel size ÷ diameter in pixels
1
2