Capture an image in
Fluoro
and name it
shim2
.
10. Correct the shim1 file and unwarp it (using Transform1).
11. Load the corrected and unwarped file into
MicroView
.
12. Use the Worksheet to make a note of the x value of the centre of the bebe using
MicroView’s
Region Grow tool.
13. Repeat steps 11 and 12 above (correct, unwarp & load into MicroView) for the
shim2
file. Use
the Worksheet to make a note of the x value of the centre of the bebe using
MicroView’s
Region Grow tool.
Note:
The x values must be within 2 pixels of each other (e.g. if one value is 1140, the other must be
within 1138 to 1142).
If the values
ARE NOT
within 2 pixels of each other:
14. Place shims behind the elevation rail (loosen stage bolts, etc.). Tighten the screws to a torque
setting of 25 in/lbs.
Note:
One pixel is approximately .002" in shims. If the x value of the top bebe in
Fluoro
is less than the x
value of the bottom bebe, the shim goes on the bottom of the stage. If the x value of the top bebe
is higher, the shim goes at the top of the stage.
The shims are placed close to the
mounting bolts. Larger adjustments
(more than 0.004") will require shims at
both the top
and
the bottom to allow the
stage to tip once it is tightened (e.g.
large shim on the bottom and small shim
on the top). This is because flat surfaces
are bolted together.
Shims
Stage Outer
Bracket
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...