Correct the Image:
35. Correct the captured image as
follows:
-
access the
Calibration Tools
program (located in
C:\Program Files\GE Medical
Systems\eXplore
Utilities\Calibration)
-
click the
Correct
tab
-
in the VFF to correct field, find
the newly captured image (e.g.
capture
)
-
use the
Browse
buttons to find
the files called
bright
and
dark
-
click the
Correct
button (the
corrected file is named
corrected.vff
and it is stored in
the same folder in which the
captured imaged was stored)
Unwarp the Image:
36. Click the
Unwarp
tab in
Calibration Tools
and complete the fields as follows:
-
in the VFF to unwarp: field use the
Browse
button to find the file D:\calib\corrected.vff
-
in the Transform grid:
field, use the
Browse
button to find the
D:\Transform1 file.
37. Click the
Unwarp
button.
38. Call the file D:\zz when
prompted to name the file,
then click
OK
.
The zz file will be used as a
temporary file for unwarping.
Continue to overwrite this file
whenever unwarping.
39. Double-click on
zz
in
Windows Explorer
to load
the image into
MicroView
.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...