LSM 510 AND LSM 510 META - SETUP REQUIREMENTS
LSM 510
DuoScan
Carl Zeiss
Vibrations / Microscopes
LSM 510 META
DuoScan
2-16
B 45-0021 e
03/06
2.6 Vibrations
Vibrations under operation conditions
(with system table)
Shipping shock (LSM 510 box)
5 µm pp at 5 Hz
3 g
10 µm pp at 10 Hz
10 µm pp at 20 Hz
2.7 Microscopes
Inverted Axiovert 200 M BP or SP
Upright Axioskop 2 FS
MOT
Upright Axio Imager.Z1
Upright Axio Imager.M1
All ICS objectives from Carl Zeiss and their accessories can be
accommodated.
Z motor
DC servomotor, opto-electronically coded
Least Z interval: 50 nm (Axiovert 200 M BP or SP)
100 nm (Axioskop 2 FS
MOT
)
10 nm (Axio Imager.Z1)
25 nm (Axio Imager.M1)
Piezo Objective focus
Piezo-driven single objective drive
Max. travel 250 µm; resolution 10 nm
In the unlikely case of extreme fluctuations of the external power net or
electromagnetic radiation, the piezo crystal will vary and disturbance in
the image is visible. Note that this is not a defect and the piezo drive will
not be damaged.