1.7 Physical
Dimensions
Length (cm)
Width (cm)
Height (cm)
Weight (kg)
Large system table
150
80 78 100
Small system table
65
80
78
60
Passively damped anti-vibration table
130 100 75 137
Active anti-vibration table (NLO)
for Mai Tai Laser or Chameleon
120 140 75 200
Active anti-vibration table (NLO)
for Verdi Mira or Millenia Tsunami
Laser
180 140 75 400
Scanning Module LSM 510
25
20
25
15
Scanning Module LSM 510 META
28
27
30.5
13
Scanning Module LSM
DuoScan
40 15 13 8
Microscope
50
35
50
20
Laser Module RGB
110 70 28 95
Laser Module, UV
140
20 20 60
Laser Module V (405 nm)
66 52 22 30
Plug-in unit external laser
66 52 22 9
System Electronic Rack
110
70 58 90
Power supply for Ar (UV)
50
50
30
30
Cooling unit for Ar (UV)
80
45
50
30
Laser Module
LIVE
66 52 22 58
1.8 Environmental
Requirements
1. Operation, specified performance
T = 22 °C ± 3 °C without interruption (24 h a day
independently whether system is operated or
switched-off)
2. Operation, reduced performance
T = 10 °C to 35 °C, any conditions different from 1.
and 5.
3. Storage, less than 16 h
T = -40 °C to 55 °C
4. Storage, less than 6 h
T = -55 °C to 70 °C
5. Temperature gradient
± 0.5 °C/h
6. Warm up time
1 h, for high-precision and/or long-term measure-
ments
≥
3 h
7. Relative humidity
< 65 % at 30 °C
8. Operation altitude
max. 2000 m
9. Loss of heat
4 kW
These requirements do not include the requirements for high precision measurements. Please
refer to the Operator’s Manual of the microscope for these requirements.