LSM 510
Routine Mode
LSM 510 META
Application of Examination Methods in …
Carl Zeiss
03/06
B 45-0021 e
6-17
•
The interval to be used during the time series is set to a default value by activating a
Time
button.
The time displayed on the time buttons can be changed as follows, if required:
•
Activate a
Time
button with a click of the mouse.
•
Use the slider (arrow buttons or input box) near
Time
to set the required interval. The set time will be
displayed online on the button. Select the required time unit by clicking on the relevant
Unit
button.
You can assign different times to all the six
Time
buttons.
•
Changing to another interval is possible by activating another
Time
button during a current time
series.
If the interval change was made via trigger key in the imported examination method, the assigned
Time
button will be activated on pressing the displayed trigger key.
(3) Marker
panel
The setting of a marker permits information about
the moment in the current time series and any
required comment to be assigned to the current
scan. The time indication is set automatically, while
comments must be defined before.
The markers (red squares) are visible in the Image
window if the
Gallery
button (
Display
toolbar) is
activated.
On storage of the image, all the markers, including
the time indication and the comments, are stored
along with the image contents.
The following functions are available:
Set 1-7
button
Setting of a marker during the scan procedure.
Edit Text
input box (1-7)
Entry of the comments for the marker.
Trigger in
field (1-7)
Display of trigger keys (1-4) used to set the marker.
Trigger out
field (1-7)
Display of trigger keys (1-4) for the out signal.
•
Click once in the edit text box of the required marker. Enter the comments via the keyboard. Then
click outside the editing box to close this box.
•
Clicking on one of the
Set 1
to
7
marker buttons will set a marker for the current scan.
Fig. 6-23
Marker panel