OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Acquire Menu
LSM 510 META
DuoScan
4-52
B 45-0021 e
03/06
4.5.3 Configuration
Control
The setting of the beam path for the scanning procedure, i.e. the definition of channels (single detector,
META detector) and tracks and the setting of the Acousto-Optical Tunable Filters (AOTF) of the various
laser lines is performed in the
Configuration Control
window.
A track is:
- a set of parameters for the detection channels and for illumination (wavelength and intensity)
- scanned simultaneously and identified and handled by the system with one name
The
Configuration Control
window has a different appearance, depending on which selection button
has been activated (
Channel
,
Lambda
Mode or Online Fingerprinting Mode
).
The
Lambda
Mode and Online Fingerprinting Mode
are only available if the system contains the
META detector. Only in the
Channel Mode
a subordinate toolbar exist which contains a
Single Track
,
Multi Track
or
Ratio
button. In the
Channel
mode use the
Single Track
and
Multi Track
buttons to
toggle between the two image acquisition modes
single tracking
and
multitracking
. In addition, you
can activate the
Ratio
button for the activation of up to two
Ratio
channels.
Performed settings can be stored as
Track Configurations
for
single tracking
. The number of
traditional channels to be defined in one track is limited to 8 (incl. monitor diode, transmission and ratio
channels), depending on the configuration. The number of channels to be defined in one track for
LSM 510 META
DuoScan
systems is 8 (including monitor diode and transmission detector). Furthermore,
2 ratio channels can be used.
In case the number of available channels is not sufficient for the scanning procedure, further tracks can
be added and configured. The combination of these tracks can also be stored as
Recording
Configurations
for
multitracking
. A recording configuration may contain the maximum of 4 tracks.
Regardless of the number of included tracks, the maximum of 8 channels (incl. monitor diode,
transmission and ratio channels) can be used in a recording configuration in
multitracking
.
If several tracks have been activated, they are processed one after the other during the scan procedure.
If the maximum number of channels to be used in a
Single Track
or a
Multi Track
has already been
achieved, it is no longer possible to add further channels or tracks.
If a second track or further tracks are used, the scan parameters can be changed as required. This avoids
"cross-talk" from one channel to another when different tracks are used.