OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Display and Analysis of Images
LSM 510 META
DuoScan
4-328
B 45-0021 e
03/06
(4)
Curve of tp measurement mode in 3D display
This function is performed in the same way as in the 2D display mode.
Before determination of the tp bearing portion, individual peaks (noise, steep slopes) must be eliminated.
The
Median
filter or a
3x3
longpass filter can be used for this purpose.
Shifting the two cursor crosses permits two bearing portions to be given in percent (e.g. Smr1 = 10 %;
Smr2 = 90 %) as default values for which the height difference Rdc is determined automatically.
(5)
Grad. Histo measurement mode in 3D display
This function is performed in the same way as in the 2D display mode.
(6)
Roughness measurement mode in 3D display
This function is performed in the same way as in the 2D display mode.
Fig. 4-308
Image Display window, Topography display: 3D – Curve of tp