LSM 510
ANNEX
LSM 510 META
Monitor Diode
Carl Zeiss
03/06
B 45-0021 e
8-23
Amplifier Gain
The diode signal is set to the right range (gray level
between 50 and 200 - 8 bit image / 750 and
3500 - 12 bit image) with the help of gray filters
and amplifier gain. The use of a lower filter density
should be prioritized against the use of a high gain
value.
The value of the amplifier gain of both channels
(PMT and diode) should be set to one, if possible.
Because of an increasing amplifier noise, parallel to
the gain factor, a gain value of more than 2 should
be avoided. The most important thing is to avoid
pixels below the zero level and beyond the
maximum range respectively.
Fig. 8-9
Scan Control window