OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Maintain Menu
LSM 510 META
DuoScan
4-226
B 45-0021 e
03/06
4.10 Maintain
Menu
The
Maintain
menu contains additional modules to check and guarantee the interference-free operation
of all the software and hardware components of the LSM 510. The LSM 510 and LSM 510 META can be
equipped with an additional scanner, the LSM
DuoScan
. For optimal results the scan fields of the two
scanners must be aligned. This procedure is described in chapter 4.8.4.1
•
In the
Main
menu toolbar, click on
Maintain
. This opens another subordinate toolbar in the
Main
menu.
4.10.1 Scanner
The
Scanner
function is used for scanner
calibration.
The following type of calibrations are available:
−
Electrical calibration with Speed 1-10
(unidirectional / bidirectional)
−
Electrical calibration with Speed 11-13
(unidirectional)
−
Optical calibration with Speed 11-13
(bidirectional)
4.10.1.1
Electrical Calibration
Electrical Calibration with Speed 1-10 can be performed unidirectional / bidirectional, but with
Speed 11-13 only unidirectional
(1) Preliminary
notes
The electrical calibration has to be performed every 2-3 months. For electrical calibration no laser
scanning is performed and for that reason no calibration sample is needed.
Fig. 4-223
Maintain menu
Fig. 4-224
Scanner Calibration window