LSM 510
LSM 510 AND LSM 510 META - SETUP REQUIREMENTS
LSM 510 META
Scanning Module
Carl Zeiss
03/06
B 45-0021 e
2-17
2.8
Scanning Module LSM 510 and LSM 510 META
Scanners
2 individually driven galvanometric scanners
Scanning speed
Up to ~5 frames/sec (512
×
512 pixels)
Field resolution
Max. 2048
×
2048 pixels (individually adjustable for each axis)
Field of view
10
×
10 mm² with a 1.25
×
objective
Zoom
1
×
... 40
×
, continuous control
Channels
a) Up to 4 confocal reflection/fluorescence channels (PMT) simultaneously
or
b) 2 confocal reflection/fluorescence channels (PMT) and 1 META detector
1 transmitted light channel (PMT) and 2 NDD
or
3 - 4 NDD (Non descanned detectors, PMT)
1 reference monitor diode
Fiber-optic adaptation of external detectors
Dynamic range
12-bit DAC for each detection channel
Pinholes
4 individual variable pinholes (one per confocal channel or META detector)
Computer controlled automatic alignment
2.9
Scanning Module LSM
DuoScan
Scanners
2 individually driven galvanometric scanners
Scanning speed
Up to ~5 frames/sec (512
×
512 pixels)
Field of view
10
×
10 mm² with a 1.25
×
objective