OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Macro Menu
LSM 510 META
DuoScan
4-202
B 45-0021 e
03/06
The other sections contain offsets for the individual tiles.
The section name has the format [Tile(xxx,yyy)] where xxx is the zero based horizontal index of the tile
and yyy is the zero based vertical index of the tile.
Entries:
TileOffsetX
TileOffsetY
The horizontal and vertical offset of the left-top corner of the tile relative to the
left-top corner of the result topography in pixels. These offset values have been
calculated with cross correlation from the original acquired stacks.
TileOffsetZ
The height offset for the data in the tile in meters. Note that the tile data
already contain height corrected data and the macro writes always the value 0.
By a multiple manual correction already manually corrected data could be more
easily distinguished from uncorrected data. The value can be modified in the file
and the
Load
command will use this value to shift the tiles in z-direction.
(6)
Loading Multiple Stack files
Multiple Stack
scan *.inf files can be loaded in the
StitchArt plus
software.