OPERATION IN EXPERT MODE
LSM 510 META
Acquire Menu
Carl Zeiss
03/06
B 45-0021 e
4-111
Set 1-7
button
Setting of a marker during the scan procedure.
Description
input box (1-7)
Entry of the comments for the marker.
Trigger in
list box (1-7)
Selection of the trigger key (1-4) with which the marker is to be set.
Trigger out
list box (1-7)
Selection of the trigger key (1-4) for the out signal.
Setting marker for the current scan
•
A marker for the current scan is set by clicking on one of the
Set 1
to
7
marker buttons.
The assignment of any required comment for the marker must be performed as follows:
•
Click in the
Edit Text
box of the required marker key (e.g.:
Set 1
) to open the editing box.
•
Enter the comments via the keyboard. Then click outside the editing box to close this box again.
You can assign comments of any required length to all the seven
Set
buttons and store this assignment
as a combination of marker keys.
•
Enter a name in the Marker list box and click on
Store
to store the combination.
If required, a combination of markers can be activated again quickly.
•
Open the Marker list box with a click on the arrow button and select the required combination with a
click of the mouse.
•
Then click on the
Apply
button to activate the combination. The relevant comments are displayed in
the
Edit Text
boxes of the
Set
buttons.
Combinations which are no longer required can be deleted.
•
Open the Marker list box and select the required combination.
•
Click on the
Delete
button. The combination will be removed.
The marker buttons can also be activated via keys 1 to 4 of the Trigger Control.
•
Click on the required
Set
button.
•
Open the
Trigger in
list box with a click on the arrow button.
•
Select one of the trigger keys 1 to 4 (e.g.
Trigger4
).
It is also possible to trigger an out signal via Trigger Control.
•
Click on the required
Set
button.
•
Open the
Trigger out
list box with a click on the arrow button.
•
Select one of the trigger keys 1 to 4 (e.g.
Trigger4
).
In this example, the relevant
Set
button is activated on pressing key
4
of the Trigger Control, a marker is
set in the Scan and an out signal given at the same time.