LSM 510
ANNEX
LSM 510 META
Specification of Trigger-Interface LSM 510 / LSM 510 META
Carl Zeiss
03/06
B 45-0021 e
8-15
8.6
Specifications of Trigger-Interface LSM 510 / LSM 510 META
Application:
With the LSM 510 / LSM 510 META you can control various actions externally using Trigger-In or force
external devices to work at a defined time depending on an action using Trigger-Out during time series.
These actions are: Scan-Start / Stop, Bleach, Change of Scan-Interval, end of a countdown, set marker
into image or even a mouse-click on a button.
Interface:
−
User Port Adapter on the back of the Real Time Electronics in the System Electronic Rack
−
Connector 2x
15-point
high density D-Type plug, 1x coax with outer shield
(Triax Lemosa Serie 00, EPL.00.650NLN)
Number:
−
8x signal IN (all able to generate an interrupt)
−
18x signal OUT (10 synchronous, 8 asynchronous relative scan position)
−
1x clock
Connector:
Coax: Triax Lemosa Serie 00, EPL.00.650NLN
Pin
Signal name
Description
Signal PCLK_Out
PixelClock
Output
Signal shield
GND
Signal ground
Outer shield
Shield
Chassis ground