LSM 510 AND LSM 510 META - SETUP REQUIREMENTS
LSM 510
DuoScan
Carl Zeiss
Power Requirements
LSM 510 META
DuoScan
2-10
B 45-0021 e
03/06
2.2 Power
Requirements
The LSM 510 and LSM 510 META comes with a mains power supply cord and plug,
either CEE red (3/N/PE 400/230V/16A), or NEMA L 14-30P (2/N/Ground 120/240V/30A), and
with the matching mains socket outlet.
A ground wire (AWG10 green/yellow) is supplied because it is necessary to ground the
system. The connecting part on both ends of the cable is a cable eye with 8 mm inner
diameter.
A suitable grounding point must be installed in the room. For systems (220 ... 240 V AC)
equipped with X-Cite 120 the mains socket outlet must be equipped with a fuse having
minimum tripping characteristic C according to IEC/EN 60898.
Line voltage
220 … 240 V AC (±10 %)
100 … 125 V AC (±10 %)
Line frequency
50...60 Hz
50...60 Hz
LSM incl. VIS laser
−
Max. current
−
Power
3 phases at 16 A
Phase 1 = 1.9 kVA max.
Phase 2 = 1.5 kVA max.
Phase 3 = 2.6 kVA max.
2 phases at 25 A
Phase 1 = 3.2 kVA max.
Phase 2 = 2.8 kVA max.
−
Power consumption 5000 VA max.
5000 VA max.
Argon UV laser
−
Line Voltage
208...240 V AC
(±10 %) 50 / 60 Hz
208...240 VAC
(±10 %) 50 / 60 Hz
−
Max. current
1 phase at 63 A
Note: For Line Voltage 220 V the
connector and power plug are rated
for 63 Amps, However wiring and
fuse should be rated for 32 Amps.
1 phase at:
208 V: 34 Amps
230 V: 31 Amps
240 V: 29 Amps
−
Power consumption 7000 VA max.
7000 VA max.
Class of protection
I
I
Type of protection
IP 20
IP 20
Overvoltage category
II
II
Pollution degree
2
2